Plasma autoclave HPS50 is distributed by Nihophawa – one of the leading brands of sterilization equipment in Vietnam. With over 11 years of experience, we will help you choose the best product as required.
Model
HPS-30
HPS-50
Sterilization subsstance
Hydrogen peroxide, 1 cycle/box
Sterilization device & time
Smart mode
40 minutes
Standard mode
48 minutes
Special mode
56 minutes
Cycle temperature
45 °C – 60 °C
SAL (Sterility assurance level)
10-6 (Bacillus)
Auxiliary product
Oxygen and steam
Sterilization chamber
Shape
Rectangle
Material
Alluminum
Capacity
30 L
50 L
Dimensions
Overall (WxHxD) mm
594 x 525 x 605
594 x 525 x 890
Chamber (WxHxD) mm
340 x 240 x 395
338 x 236 x 660
Weight (kg)
90
110
Condition
32-bit microprocessor
Information on display
Cycle 2.7 inch (OLED), thermal printer
Power
220-230 VAC, 50/60Hz, Max 2.7 kW
220-230 VAC, 50/60Hz, Max 3.0 kW
Requirement of device installation distance
Front: 100cm
Right, left and back: 5 cm
Transporting cart system can be selected for easy movement of the device
Printer
Built-in thermal printer
Dimensions of transporting cart (WxDxH) mm
613 x 665 x 850
613 x 880 x 850
Do not use the plasma autoclave HPS50 to sterilize the following materials and objects:
Products and materials that the manufacturer prohibits from being used for gas autoclave
Sterilization objects in powder or liquid form.
Sterilization of items in normal saline form.
Sterilization objects contain materials that are able to remove solvents.
Objects containing fibers such as cotton. cardboard, linen, towels, gauze, sponge or wool.