• Autoclave AS50
  • Autoclave AS75
  • Autoclave AS100

Vacuum technology, modern design.

Nihophawa autoclave uses saturated steam at high temperature to sterilize packaged and unpackaged medical instruments, surgical instruments, metals, test instruments, warming and dissolving culture medium for hospitals, clinics, research institutes, laboratories.

REQUEST A QUOTE
Nồi hấp AS50

NIHOPHAWA AUTOCLAVE AS50 – AS100

Nihophawa autoclave 55 – 110 L owns modern vacuum technology, modern design, easy to operate and configure based on purpose of use, suitable for clinics, hospitals, medical centers

Nồi hấp Nihophawa 55L
  • Model: NIHOPHAWA-AS50
  • Brand: NIHOPHAWA
  • Manufacturer: Hong Phat tech Co., Ltd

Configuration

  • Main unit: 01 set
  • Integrated vacuum pump: 01 set
  • Integrated heat exchanger: 01 set

Standard accessories:

  • Sterilization basket: 01 pc

Power cord: 01 pc

Nồi hấp Nihophawa 55L

Technical specifications of Nihophawa autoclave AS50 – AS100

Nồi hấp AS50
  • Microprocessor controller integrated with pre-installed software, parameters and user interface are displayed via touch keypad and LED touchscreen
  • Sterilization pressure range: : 0 kgf/cm² to 2.3 kgf/cm²
  • Power supply: 220 VAC + 10%/50 Hz
  • Sterilization chamber: made of SUS 304 or SUS 316 stainless steel
  • Timer mode in real-time and calender
  • Temperature sensor (Optional)
  • Pre-vacuum mode (Optional)
  • Vacuum sterilization (Optional)
  • Water supply system: Automatic

The customized program depends on the need of the customer

There are 10 standard programs

  • Program 1: Heating -Warming
  • Program 2: Dissolving mode (Heating-Warming)
  • Program 3: Sterilization of liquid in the opening container
  • Program 4: Sterilization of liquid in the closing container
  • Program 5: Sterilization of tools in the opening form, gloves, plastics
  • Program 6: Sterilization of tools in the closing form, gloves, heat-resistant plastics
  • Program 7: Sterilization of fabric,100% PE (Polyester) clothes and stainless steel tools.

• Program 8: Sterilization of fabric, 35/65 Cotton (Tixi) clothes and stainless steel tools

• Program 9: Sterilization of fabric, 65/35 Cotton (CVC) clothes and stainless steel tools.

. Program 10: Sterilize fabric, 100% cotton clothes and stainless steel tools

Other specifications:

  • Sterilization temperature range: 100°C ÷ 136°C (± 1°C)
  • Heating temperature range: 40°C ÷ 60°C (± 1°C)
  • Temperature range for dissolving mode: 40°C ÷ 99°C (± 1°C)
  • Temperature range for digital protection: 100°C ÷ 140°C (± 1°C)
  • Sterilization time range: 0 min ÷ 100 hours 0 min (with an increment of 1 min)

Drying time range: 0 min ÷ 100 hours 0 min (with an increment of 1 min)

LED display

  • Display installed parameters: Temperature, sterilization time, drying time, selected program, error messages for each stage of operating process.
  • Display descriptive diagram: sterilization cycle via single LED and digital LED
  • Display srerilization and drying parameters, and warning via digital LED screen.
  • Single LED displays weekly calendar from Monday to Sunday
  • Digital LED display: 3 displays
  • LED display 1: Display real-time clock
  • LED display 2: Display installed parameters and values in real-time.
  • LED display 3: Display the installed time parameters
  • Display installed drying time
  • Display the remaining drying time during the device operation
  • Display time value for device operation in real-time
  • Single LEDs display the operation procedures for each installed program.
  • Values and parameters: supplying water, removing gas, heating, sterilizing, rinsing, drying, ending with diagrams and language Signaling by sound at the end of the cycle
  • Display warnings

Printer (Optional): Automatically print parameters during the sterilization process.

  • Name of User
  • Address of User
  • Product’s model
  • Product’s Serial number
  • Starting time of operation
  • Time of performing sterilization cycles, temperature, pressure of each cycle
  • Ending time of sterilization program
  • Average temperature coefficient during sterilization (Ft).
  • Average pressure coefficient during sterilization (Ft).
  • Sterilization result: Pass or fail.

Signature position of operator

Warning and safety protection functions.

  • Safety valve: operates at pressure of 2.7 kgf/cm².
  • Digital overheat protection, from 100°C to 140°C and automatically discharging the exhaust valve to ensure safety for device and user.
  • Automatic error reporting.
  • After 15 minutes when there is insufficient or no water supplied to the device’s water cycle
  • Lack of water during the sterilization process
  • When the temperature sensor is faulty and automatically stops working.
  • When the resistor breaks and power supply is automatically disconnected.
  • Emergency stop button (in case the automatic valve fails, requiring the user’s intervention).

Anti-leakage of electricity through the anti-shock circuit breaker